MAUDE Adverse Event Report: COOK ENDOSCOPY CAPTURA SERRATED FORCEPS WITH SPIKE; FCL, FORCEPS, BIOPSY, NON-ELECTRIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY CAPTURA SERRATED FORCEPS WITH SPIKE; FCL, FORCEPS, BIOPSY, NON-ELECTRIC Back to Search Results
Catalog Number DBF-2.4SN-230SP-20-S
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation evaluations: our laboratory evaluation of the product said to be involved confirmed the report.Two (2) devices were returned for evaluation.Both devices were evaluated by performing a functional test.During the functional test on both devices, the handles were manipulated and the forceps cups would open, but they would not close.Multiple attempts to close the devices were made.The devices were sent back to the supplier for further evaluation.The supplier provided the following information: two devices from the reported event lot were returned in a zip type bag with proof of decontamination.The devices were tested for "would not close".During functional testing, it was confirmed that the devices would open but would not close when manipulated using the handle.Upon disassembly of the device tip, it was noted that the solder joint was broken on each device.The reported was confirmed for both devices.The device history records were reviewed and the date of manufacture was february 2017.The order had one device rejected for solder joint defects in manufacturing and/or final quality control.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: the customer experienced issue of "would not open/close" was confirmed; the device has a butt joint that has broken at the solder connection.The root cause is a butt joint with a broken control wire / link wire solder joint.The supplier is currently working on improvements to create a more robust soldering process to prevent solder joints from breaking.Instructions for use states: "beginning at the handle and moving toward the cups, uncoil the forceps making sure not to stretch the cable.Open and close the cups to verify smooth handle operation and appropriate cup action.Become familiar with the amount of handle movement required to operate the cups.If any irregularities are noted, do not use.Note: exercising the handle while the forcep is coiled may result in damage to the performance characteristics of the forceps." prior to distribution, all captura biopsy forceps are subjected to a visual inspection and functional test to ensure proper workability.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During an endoscopic procedure, the physician used two (2) cook captura serrated forceps with spike.The cups would not open nor close.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURA SERRATED FORCEPS WITH SPIKE
Common Device Name
FCL, FORCEPS, BIOPSY, NON-ELECTRIC
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key6585463
Report Number1037905-2017-00300
Device Sequence Number10918592
Product Code FCL
UDI-Device Identifier00827002561328
UDI-Public(01)00827002561328(17)200215(10)W3831969
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberDBF-2.4SN-230SP-20-S
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2017
Initial Report FDA Received Date05/23/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-