| Model Number |
10218 |
| Medical Device Problem Code |
Air Leak (1008)
|
| Health Effect - Clinical Code |
Death (1802)
|
| Date of Event |
04/20/2017
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that during the delivery of cardioplegia with this cannula, the customer noted entrainment of air.The device was replaced with another cannula from the same lot, but the same issue was noted.The customer later noted the upper connection port on the cannula could easily be separated from the tip.The patient passed away one day after the case.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|