|
Investigation - evaluation: a review of complaint history, device history record, drawing, instructions for use, manufacturing instructions, quality control, specifications, dimensional verification, and visual inspection of the actual device were completed during this investigation.The device was returned in two pieces.The distal tip was returned severely elongated and separated from the proximal end.The coil diameter measured within specification but the mandril wire was severely elongated.One potential reason that the wire guide does not withstand forces during insertion, manipulation or withdrawal is excessive friction forces.Another cause is that the wire guide is withdrawn through the needle.The warning label on the wire guide holder cautions the user to not pull the distal spring coil portion of the wire guide through the needle tip.This is because pulling the coil through the needle may cause the wire guide to elongate, break or become stuck.Based on the available information it is not clear if the leading cause to this event might be associated with the medical procedure / intervention / user technique inaccuracy or an eventual malfunction of the device.While no definitive cause can be attributed to this failure, it is feasible that coils became stuck on the needle bevel during a manipulation procedure that involved some reversal of the coil through the needle tip.However, lacking evidence of this, the root cause will be listed as inconclusive.Appropriate personnel have been notified to monitor for similar complaints.Per the risk assessment no further action is required.
|