• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT LABORATORIES PRISM CHAGAS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT LABORATORIES PRISM CHAGAS Back to Search Results
Catalog Number 07K35-68
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Additional information suspect medical device; concomitant: abbott esa chagas; list # 08l34-68; lot # a00001300.
 
Event or Problem Description
The customer reported increase in (b)(6) prism chagas and prism hcv results in (b)(6)2017.Of 14 rr chagas samples 7 were (b)(6).Supplemental/confirmatory testing was performed with ortho chagas (2 (b)(6), 5 not tested); ortho hcv all 7 (b)(6); abbott chagas esa (6 (b)(6), one not tested).There was no reported impact to donor/patient management.There was no additional donor information provided.
 
Additional Manufacturer Narrative
Evaluation of complaint data for the lots identified normal complaint activity.A review of the manufacturing documentation did not identify any issues associated with the customer observation.Labeling was reviewed and found to adequately address the issue.A review of the prism metrics field data indicates that the initial and repeat reactive rates for the abbott prism chagas lot 68048m500 which is less than the package insert's upper 95% confidence intervals.Labeling claims were met for specificity; there is no product deficiency.A malfunction was not identified.Based on the information within the complaint record, the device met performance specifications and performed as intended at the customer site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISM CHAGAS
Common Device Name
CHAGAS
Manufacturer (Section D)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064 3500
MDR Report Key6587614
Report Number1415939-2017-00114
Device Sequence Number11007534
Product Code MIU
UDI-Device Identifier00380740062941
UDI-Public(01)00380740062941(17)170625(10)68048M500
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 06/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/25/2017
Device Catalogue Number07K35-68
Device Lot Number68048M500
Was Device Available for Evaluation? Yes
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 05/04/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/24/2017
Supplement Report FDA Received Date06/07/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LOT #72105M500 AND 70080M500,; SERIAL # (B)(4), PRISM HCV, LIST # 06D18-68,
-
-