| Catalog Number |
07K35-68 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Additional information suspect medical device; concomitant: abbott esa chagas; list # 08l34-68; lot # a00001300.
|
| |
|
Event or Problem Description
|
|
The customer reported increase in (b)(6) prism chagas and prism hcv results in (b)(6)2017.Of 14 rr chagas samples 7 were (b)(6).Supplemental/confirmatory testing was performed with ortho chagas (2 (b)(6), 5 not tested); ortho hcv all 7 (b)(6); abbott chagas esa (6 (b)(6), one not tested).There was no reported impact to donor/patient management.There was no additional donor information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of complaint data for the lots identified normal complaint activity.A review of the manufacturing documentation did not identify any issues associated with the customer observation.Labeling was reviewed and found to adequately address the issue.A review of the prism metrics field data indicates that the initial and repeat reactive rates for the abbott prism chagas lot 68048m500 which is less than the package insert's upper 95% confidence intervals.Labeling claims were met for specificity; there is no product deficiency.A malfunction was not identified.Based on the information within the complaint record, the device met performance specifications and performed as intended at the customer site.
|
| |
|