| Model Number |
B2-70072 |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
05/16/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The affected iv set is being sent to the contract supplier for investigation.Zyno medical has sent a quality alert to the contract supplier on 05/19/2017.
|
| |
|
Event or Problem Description
|
|
The user reported an iv set leaking issue.A hole was created in the tubing when the pinch clamp was released."the iv set that the nurses were testing on (demonstration) in which she tried to disengage the pinch clamp and released it, but during the release of the pinch clamp it caused a small hole in the tubing." no patient was involved in this case.
|
| |
|
Additional Manufacturer Narrative
|
|
The user reported issue was confirmed.Zyno medical has received the testing report from the contract supplier on 07/17/2017.The root cause of the hole of the tubing was identified as equipment damage.The acrylic guard, which covered the calibrated ruler that used to cut the tubing, had a crack and could potentially cause damage in the tubing.Details of the testing results can be found in the attached report.
|
| |
|
Event or Problem Description
|
|
This is a follow-up for the initially filed mdr (3006575795-2017-00130).
|
| |
|