Model Number B2-70072 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
Not Applicable (3189)
|
Event Date 05/16/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
The affected iv set is being sent to the contract supplier for investigation.Zyno medical has sent a quality alert to the contract supplier on 05/19/2017.
|
|
Event Description
|
The user reported an iv set leaking issue.A hole was created in the tubing when the pinch clamp was released."the iv set that the nurses were testing on (demonstration) in which she tried to disengage the pinch clamp and released it, but during the release of the pinch clamp it caused a small hole in the tubing." no patient was involved in this case.
|
|
Manufacturer Narrative
|
The user reported issue was confirmed.Zyno medical has received the testing report from the contract supplier on 07/17/2017.The root cause of the hole of the tubing was identified as equipment damage.The acrylic guard, which covered the calibrated ruler that used to cut the tubing, had a crack and could potentially cause damage in the tubing.Details of the testing results can be found in the attached report.
|
|
Event Description
|
This is a follow-up for the initially filed mdr (3006575795-2017-00130).
|
|
Search Alerts/Recalls
|