• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAUSCH + LOMB OPTIMIZED STABILITY VACUUM PACK; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAUSCH + LOMB OPTIMIZED STABILITY VACUUM PACK; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number BL5112
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/24/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product is not available for evaluation.Sterilization and lot history records were reviewed and no exceptions were found.Report 1 of 3.
 
Event or Problem Description
The user facility reported observing glittery metallic particles in the incision.A few of the particles were observed on the iris in close proximity to the main incision.In one case the particles were observed on the conjunctiva in the area of the main incision.There was no injury or impact to the patients.
 
Additional Manufacturer Narrative
Based on the results of our root cause investigation, it has been determined that the reported particulate is not related to any damage or malfunction of the product.Our investigation has concluded that the observed particles are originating from the metallic cap assembled to the needle wrench which when packaged in the same pouch with the silicone irrigation sleeve results in microscopic particles being transferred from the needle wrench cap to the irrigation sleeve.All products have been found to comply with specifications and function as intended.Additionally, there have been no reports of post-surgical adverse reaction.However, due to increased sightings of microscopic particles and our ongoing continuous improvement activities, we have taken measures to identify an internal cleaning process related to needle wrench assembly to address the reported events.Validation activities have been completed and the process has been implemented.In addition, we are reviewing our component specifications related to particulate matter and will revise specifications, where necessary; to better align with customer expectations.As a result of this additional information, it has been determined that the issue documented in this report does not meet the criteria of an reportable malfunction and the complaint record will be revised accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIMIZED STABILITY VACUUM PACK
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
BAUSCH + LOMB
rochester NY 14609
MDR Report Key6587923
Report Number0001920664-2017-00215
Device Sequence Number13769780
Product Code HQC
UDI-Device Identifier10757770057082
UDI-Public(01)10757770057082(17)171205
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K063331
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 04/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/05/2017
Device Model NumberBL5112
Device Lot NumberV7305
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 04/25/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/24/2017
Supplement Report FDA Received Date06/08/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-