MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-40
Device Problems Kinked (1339); Difficult to Advance (2920); Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/15/2017
Event Type  malfunction  
Manufacturer Narrative
Other relevant components include: product id 8780 lot# serial# (b)(4) implanted: (b)(6) 2017.Product type catheter (b)(4).A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare professional via a manufacturer representative regarding a patient receiving gablofen with an unknown dose and concentration via an implantable for intractable spasticity.It was reported the healthcare professional could not advance the catheter.It was noted that the catheter was not able to be advanced and kinked.It was noted that the healthcare professional (hcp) tried to advance the catheter many times, and the hcp believed that catheter was kinking more easily that they had seen before.It was noted that the hcp pulled the needle out and then advanced the catheter without the needle in place with a twisting motion.It was noted the patient had good cerebrospinal fluid (csf) backflow once it was placed.It was stated that the patient has no abnormal anatomy or hardware.No symptoms were mentioned as it was a new pump and catheter.Event date was noted as (b)(6) 2017.No further complications were reported/anticipated.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key6588622
MDR Text Key75902155
Report Number3004209178-2017-10974
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169508156
UDI-Public00643169508156
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/24/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/14/2018
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/15/2017
Date Device Manufactured03/21/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age65 YR
-
-