| Model Number |
980 |
| Medical Device Problem Code |
Device Inoperable (1663)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Information (3190)
|
| Date of Event |
04/20/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that, a 980 ventilator failed the power on self test (post) and generated an inoperable error message.The ventilator was not in use on a patient at the time of the reported event.The service engineer (se) inspected the device and verified the reported issue.The se replaced the universal serial bus (usb).The se performed calibrations and extended self-testing on the device and all tests passed.
|
| |
|