| Model Number |
2200P-47E |
| Medical Device Problem Codes |
Device Inoperable (1663); Device Displays Incorrect Message (2591); Material Distortion (2977); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Codes |
Therapeutic Response, Decreased (2271); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/24/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Explanted device not yet returned.
|
| |
|
Event or Problem Description
|
|
This subject is a participant in the (b)(6) clinical trial.This subject was originally in the control arm of the trial and underwent implantation of the active maestro rechargeable system on (b)(6) 2014.Impedance issues were reported related to the implanted anterior lead (reference mdr3005025697-2017-00001 for information on that component).During a revision procedure to replace the anterior lead, it was discovered that the anterior lead and posterior lead were twisted around each other in the pocket of the rechargeable neuroregulator.As a result, the posterior lead was also replaced.The revision procedure was successfully completed without adverse impact to the subject.Explanted device components have not yet been received by enteromedics.Reference mdr3005025697-2017-00001 for report related to the anterior lead.
|
| |
|
Event or Problem Description
|
|
This subject is a participant in the (b)(6) clinical trial.This subject was originally in the control arm of the trial and underwent implantation of the active maestro rechargeable system on (b)(6) 2014.Impedance issues were reported related to the implanted anterior lead (reference mdr 3005025697-2017-00001 for information on that component).During a revision procedure to replace the anterior lead, it was discovered that the anterior lead and posterior lead were twisted around each other in the pocket of the rechargeable neuroregulator.As a result, the posterior lead was also replaced.The revision procedure was successfully completed without adverse impact to the subject.Explanted device components have not yet been received by enteromedics.Reference mdr 3005025697-2017-00001 for report related to the anterior lead.Supplement 001: explanted device component received by enteromedics on june 1, 2017.
|
| |
|
Search Alerts/Recalls
|