| Catalog Number |
42412 |
| Medical Device Problem Code |
Occlusion Within Device (1423)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/25/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product was unavailable for return.Therefore an evaluation of the device performance was not possible.A review of the manufacturing records showed no anomalies.All valves are 100% tested at the time of manufacture.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported to medtronic neurosurgery that the flushing of the device was unsuccessful prior to use.According to the report, the device was never implanted and was replaced.The new device flushed without any issues.Reportedly, there was no injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned valve was patent.It met the requirements for reflux, pressure-flow, preimplantation, and leak testing.Therefore the c onditions of the complaint could not be duplicated by laboratory personnel.There was crystalline debris was observed within the interior and exterior of the valve.Debris within the valve may hold pressure-flow controlling mechanisms open affecting the flow of fluid through the valve.The instructions for use (ifu) that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization, or other debris.¿ the instructions for use that accompany the device indicate that to perform the patency check: a.Attach 16-gauge blunt needle adapter, tubing, and syringe to the valve inlet.B.Pre-fill valve and tubing.C.Occlude inlet tubing.D.Depress dome.If fluid flows out of the outlet connector, the valve is patent.A review of the manufacturing records showed no anomalies.All valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter updated.If information is provided in the future, a supplemental report will be issued.
|
| |
|