MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) MEDTRONIC FUSION COMPACT NAVIGATION SYSTEM; EAR NOSE AND THROAT STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) MEDTRONIC FUSION COMPACT NAVIGATION SYSTEM; EAR NOSE AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION COMPACT
Device Problems Communication or Transmission Problem (2896); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/26/2017
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.The representative reported that the computer for the navigation system was replaced to resolve the reported issue.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The suspect computer for the navigation system was returned to the manufacturer for analysis.The computer was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
 
Event Description
A medtronic representative reported that, while in a functional endoscopic sinus surgery (fess), the navigation system emitter disconnected following patient registration.The emitter showed that it was not running on the computer.Selecting the emitter button on the navigation system restored functionality to the system.No additional information was provided.There was no reported delay to the procedure due to this issue.There was no impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDTRONIC FUSION COMPACT NAVIGATION SYSTEM
Type of Device
EAR NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key6590586
MDR Text Key75969420
Report Number1723170-2017-02158
Device Sequence Number1
Product Code PGW
UDI-Device Identifier00643169672956
UDI-Public00643169672956
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153247
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 05/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION COMPACT
Device Catalogue Number9735602
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/19/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/26/2017
Initial Date FDA Received05/25/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age34 YR
Patient Weight86
-
-