Model Number 233-01-112
Health Effect - Clinical Code Fall (1848)
Date of Event 04/28/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Second revision surgery - due to the patient falling and was not wearing his brace.He tore his medial collateral ligament (mcl) and the surgeon determined a revision would be the best course of action.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to a torn medial collateral ligament (mcl).The previous surgery and the revision detailed in this investigation occurred over 8 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported components used in the previous surgery met design and manufacturing requirements.There was a non-conforming material reports (ncmr) # (b)(4) associated with the part 233-01-112, 3d knee cemented femoral component, size 12, which documents a non-conformance that out of (b)(4) quantity lot (b)(4) items were rejected due to out of tolerance features in anterior-posterior box geometry and these parts were sent to supplier to bend within specification.After these were reworked with suitable solution, (b)(4) of the rejected items were accepted and 1b)(4) rejected item was scrapped.In total, (b)(4) out of (b)(4) items in the lot were accepted.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to torn mcl.The root cause of the torn mcl was reported as a fall; and the patient was not wearing their brace.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3DKNEE
Common Device Name
FEMUR, NONPOROUS, 3D, SIZE12, LT
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key6595887
Report Number1644408-2017-00382
Device Sequence Number10789577
Product Code MBH
UDI-Device Identifier00888912111188
UDI-Public(01)00888912111188
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K032905
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2021
Device Model Number233-01-112
Device Catalogue Number233-01-112
Device Lot Number897A1095
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/28/2017
Supplement Date Received by Manufacturer06/12/2017
Initial Report FDA Received Date05/26/2017
Supplement Report FDA Received Date07/10/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
-
-