• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MARVAO MEDICAL NEXSITE HD STEP TIP; HEMODIALYSIS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MARVAO MEDICAL NEXSITE HD STEP TIP; HEMODIALYSIS CATHETER Back to Search Results
Catalog Number NEXHD1552801
Health Effect - Clinical Code No Information (3190)
Date of Event 03/21/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was indicated that the device will not be returned for evaluation because the device remains implanted; therefore an analysis of the complaint device could not be completed.A review of the manufacturing documentation confirmed that the lot met the specification requirements.The review found no deviations or non-conformances that would have contributed to the reported complaint.Taking into consideration the evaluation conducted and the details of the complaint, this investigation was assigned the most probable root cause of an anticipated procedural complication.A complaint with a most probable root cause classification of an anticipated procedural complication indicates that a device related root cause does not apply and the complaint is due to a known effect of the procedure.
 
Event or Problem Description
(b)(4).The nexsite device was successfully placed on (b)(6) 017.On (b)(6) 2017, the patient experienced poor flow of the haemodialysis catheter that was mild in severity.No action was taken with the study device and the event resolved on the same day after treatment with alteplase.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEXSITE HD STEP TIP
Common Device Name
HEMODIALYSIS CATHETER
Manufacturer (Section D)
MARVAO MEDICAL
innovation in business centre
gmit, dublin road
galway, H91 D CH9
EI  H91 DCH9
MDR Report Key6596447
Report Number3008110587-2017-00010
Device Sequence Number16431353
Product Code MSD
UDI-Device Identifier05391525640020
UDI-Public(01)05391525640020(17)170401(10)43967
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K121933
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source study
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date04/01/2017
Device Catalogue NumberNEXHD1552801
Device Lot Number43967
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2017
Initial Report FDA Received Date05/29/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age68 YR
Patient Weight71
-
-