|
Results: the px slim was fractured approximately 35.0, 97.0, and 116.0 cm from the proximal hub.There was slight bends along the length of the catheter shaft.There was a distinct kink approximately 5.0 cm distal to the most distal fracture of the px slim.Conclusions: evaluation of the device revealed several fractures along the px slim catheter shaft and a distinct kink just distal to the most distal fracture.This kink is likely a result of the packaging tray becoming closed on the catheter, therefore, pinning the catheter in the tray.Subsequently, if the catheter is forcefully retracted while pinned, the px slim may become fractured.Further evaluation revealed there were multiple fractures on the catheter shaft.It is unlikely for the catheter to fracture in multiple locations simultaneously.Therefore, one or more of these fractures likely occurred after the initial fracture occurrence.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|