| Model Number |
48811116 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/04/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; one of the bars from the locking mechanism of the aviator plate was bent and sort of broken.The surgeon noticed the plate was after implanting the screws and when he looked under the microscope he noticed the bent, almost broken bars.Surgeon removed the screws and the plate and put another plate in.No adverse consequences to the patient and no surgical delay.Procedure was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Lot# 14c849.Method: visual inspection, device history review, complaint history review, risk assessment.Result: manufacturing files were reviewed and no anomalies were found.Conclusion: the potential root cause include:user tilts guide off axis affecting screw path, over-angulation of screwdriver which leads to unintended cantilever force, guide not properly attached to plate, guide/awl limits visibility, blocker in locked position and spring bar does not freely move as screw progresses.
|
| |
|
Event or Problem Description
|
|
It was reported that; one of the bars from the locking mechanism of the aviator plate was bent and sort of broken.The surgeon noticed the plate was after implanting the screws and when he looked under the microscope he noticed the bent, almost broken bars.Surgeon removed the screws and the plate and put another plate in.No adverse consequences to the patient and no surgical delay.Procedure was completed successfully.
|
| |
|