No report of procedure delays or cancellations.During the reported event, an employee initiated a dart® (daily air removal test) cycle when water droplets came in contact with the employee's hand causing the reported burn.The employee was not wearing proper ppe specifically gloves.The employee sought medical treatment at the emergency department and returned to work.The 16" century sterilizer operator manual (1-1) states, "warning-burn hazard: sterilizer and rack/shelves will be hot after cycle is run.Always wear protective gloves and apron (also face shield if processing liquids) when removing a processed load.Protective gloves and apron must be worn when reloading sterilizer following previous operation." a steris service technician arrived on-site, inspected the 16" century sterilizer, and identified the s2 valve in the steam manifold assembly was not operating properly.The technician identified the s2 valve seat in the assembly was seeping water droplets into the sterilizer's chamber due to condensation.The technician replaced the unit's s2 valve seat in the steam manifold assembly, tested the unit, and returned it to service.The unit was installed at the customer's location in (b)(6) 1999 and has been in use for approximately 18 years.No additional issues have been reported.
|