• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. 3DKNEE; INSERT, 3D EX SZ 6RT 11MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. 3DKNEE; INSERT, 3D EX SZ 6RT 11MM Back to Search Results
Model Number 392-11-706
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 05/02/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient's knee having dislocated; the surgeon removed and replaced the original insert.
 
Additional Manufacturer Narrative
The reason for this revision surgery was the patient's knee dislocated.The previous surgery and the revision detailed in this investigation occurred over 1 year and 1 month apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to the dislocation.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the patient's dislocation.There are multiple factors that may contribute to the event that are outside the control of djo surgical are patient activities, inadequate soft tissue, excessive range-of-motion, excessive load or trauma.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities that may have contributed to the event and hence a definitive root cause cannot be determined.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3DKNEE
Common Device Name
INSERT, 3D EX SZ 6RT 11MM
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key6603334
Report Number1644408-2017-00413
Device Sequence Number981079
Product Code OIY
UDI-Device Identifier00888912120531
UDI-Public(01)00888912120531
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K091956
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2018
Device Model Number392-11-706
Device Catalogue Number392-11-706
Device Lot Number59604334
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/02/2017
Supplement Date Received by Manufacturer06/21/2017
Initial Report FDA Received Date05/31/2017
Supplement Report FDA Received Date07/20/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
-
-