| Model Number |
E360 |
| Medical Device Problem Code |
Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the e360 ventilator oxygen supply was lost and had leaking noise.Patient involvement is unknown.Medtronic was not authorized to evaluate/service the device so the reported complaint could not be confirmed.A non-medtronic service provider checked the ventilator and verified the problem.
|
| |
|
Additional Manufacturer Narrative
|
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the e360 ventilator oxygen supply was lost and had leaking noise.The ventilator was not in use on a patient at the time of the reported event.Medtronic was not authorized to evaluate/service the device so the reported complaint could not be confirmed.A non-medtronic service provider checked the ventilator and verified that a disconnected tubing causing the noise.
|
| |
|