According to the hospital:"after almost 15 minutes of use (from the connection of the patient), the oxygenator module showed a high resistance to the blood flow, with the consequence of higher pressure drop.Since it was impossible to continue the procedure in this way, the oxygenator module has been replaced." (b)(4).
|
(b)(4).The product was investigated in the laboratory of the manufacturer.The oxygenator was cleaned and sawed off.The mats were inspected according to lv 7544.Between the mats small clots were detected.No other abnormalities were detected.Dhr review: affected product: basic lot 70109322 and packaging lot 70111067 (serial number (b)(4)).The avz 7568 from (b)(4) was reviewed on 2017-09-26.There were no references found, which are indicating a nonconformance of the product in question.Additionally a review of the weekly performance monitoring according to (b)(4) has been reviewed (dms# (b)(4)),the performance tests passed the acceptance criteria.The cause of the reported failure was determined to not be attributed to a device related malfunction.The detected clots between the mats were leading to the conclusion the clotting / micro clotting was causing the observed resistance to the blood flow.Thus a product malfunction could not be confirmed.
|