• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (GFO) LUTONIX 035 DRUG COATED BALLOON PTA CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (GFO) LUTONIX 035 DRUG COATED BALLOON PTA CATHETER Back to Search Results
Model Number 9004
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Reocclusion (1985)
Date of Event 12/07/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Analysis: the sample was not returned to the user facility; therefore, device evaluation is unable to be performed.A lot history review revealed 2 additional reocclusion complaints were associated with this lot.A review of the device history record (dhr) shows the lot was manufactured to specification.Conclusion: the actual sample was not received for evaluation.The dhr found nothing to indicate a manufacturing related cause for this event.The investigator assessed that the event was not related to the study device or procedure, but the independent reviewer assessed the event as related to the device.Based on the instructions for use (ifu), the occurrence of a reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.If additional information becomes known to the manufacturer, a supplemental report will be provided will all relevant information.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported through a clinical registry that during the index procedure, a lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheter was used to treat the target lesion located in the mid left superficial femoral artery (sfa).Approximately 14 months after the index procedure, the patient¿s mid left sfa vessel was reportedly reoccluded.A revascularization was performed and the hcp deemed it was successful.The investigator assessed that the event was not related to the study device or procedure, but the independent reviewer assessed the event as related to the device.The sample was discarded by the user facility and not available for further evaluation.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
Patient weight (b)(6).The sample was not returned to the user facility; therefore, device evaluation is unable to be performed.A lot history review revealed 2 additional reocclusion complaints were associated with this lot.A review of the device history record (dhr) shows the lot was manufactured to specification.The actual sample was not received for evaluation.The dhr found nothing to indicate a manufacturing related cause for this event.The investigator assessed that the event was not related to the study device or procedure, but the independent reviewer assessed the event as related to the device.Based on the instructions for use (ifu), the occurrence of a reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.If additional information becomes known to the manufacturer, a supplemental report will be provided will all relevant information.
 
Event or Problem Description
It was reported through a clinical registry that during the index procedure, a lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheter was used to treat the target lesion located in the mid left superficial femoral artery (sfa).Approximately 14 months after the index procedure, the patient¿s mid left sfa vessel was reportedly reoccluded.A revascularization was performed and the hcp deemed it was successful.The investigator assessed that the event was not related to the study device or procedure, but the independent reviewer assessed the event as related to the device.The sample was discarded by the user facility and not available for further evaluation.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUTONIX 035 DRUG COATED BALLOON PTA CATHETER
Common Device Name
DRUG COATED BALLOON PTA CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
MDR Report Key6604194
Report Number3006513822-2017-00109
Device Sequence Number19081317
Product Code ONU
UDI-Device Identifier00801741088667
UDI-Public(01)00801741088667(17)170716(10)GFZG1983
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P130024
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other,stu
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/16/2017
Device Model Number9004
Device Catalogue NumberLX351304100
Device Lot NumberGFZG1983
Was Device Available for Evaluation? No
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer10/02/2017
Initial Report FDA Received Date06/01/2017
Supplement Report FDA Received Date10/16/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient Weight72
-
-