MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE MAVERICK²¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE MAVERICK²¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H7493892812250
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/07/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Device evaluated by mfr: returned product consisted of a maverick 2 balloon catheter in two pieces.The shafts, balloon, tip and welds were microscopically examined.There was contrast in the inflation lumen.The balloon was tightly folded.Microscopic examination and tactile inspection revealed numerous kinks throughout the hypotube and a complete hypotube separation 43 cm from the strain relief.The hypotube fracture surfaces were ovaled, which suggests the device was kinked prior to separation.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.No other damage or issues were identified during the product analysis.Device analysis determined the condition of the returned device was not consistent with the reported information due to the additional damage found.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical/procedural factors.(b)(4).
 
Event or Problem Description
Same case as mdr id: 2134265-2017-05282.Reportable based on device analysis completed on 12may2017.It was reported that crossing difficulties were encountered.The target lesion was located in the left anterior descending artery.After a non-bsc guide wire was placed in the lesion, a 2.50 mm x 12 mm maverick²¿ balloon catheter was advanced for dilatation but failed to cross the lesion.The device was removed and another 2.50 mm x 12 mm maverick²¿ balloon catheter was advanced but also failed to cross.The device was removed and the procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed hypotube broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAVERICK²¿
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
MDR Report Key6604200
Report Number2134265-2017-05281
Device Sequence Number11015395
Product Code LOX
UDI-Device Identifier08714729369981
UDI-Public(01)08714729369981(17)20190930(10)19748140
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2019
Device Model NumberH7493892812250
Device Catalogue Number38928-1225
Device Lot Number19748140
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2017
Initial Report FDA Received Date06/01/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDE WIRE: FIELDER XT
Patient Age79 YR
Patient Weight88
-
-