|
(b)(4).Device evaluated by mfr: returned product consisted of a maverick 2 balloon catheter in two pieces.The shafts, balloon, tip and welds were microscopically examined.There was contrast in the inflation lumen.The balloon was tightly folded.Microscopic examination and tactile inspection revealed numerous kinks throughout the hypotube and a complete hypotube separation 43 cm from the strain relief.The hypotube fracture surfaces were ovaled, which suggests the device was kinked prior to separation.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.No other damage or issues were identified during the product analysis.Device analysis determined the condition of the returned device was not consistent with the reported information due to the additional damage found.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical/procedural factors.(b)(4).
|
|
Same case as mdr id: 2134265-2017-05282.Reportable based on device analysis completed on 12may2017.It was reported that crossing difficulties were encountered.The target lesion was located in the left anterior descending artery.After a non-bsc guide wire was placed in the lesion, a 2.50 mm x 12 mm maverick²¿ balloon catheter was advanced for dilatation but failed to cross the lesion.The device was removed and another 2.50 mm x 12 mm maverick²¿ balloon catheter was advanced but also failed to cross.The device was removed and the procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed hypotube broken.
|