Model Number M00560150 |
Device Problem
Material Separation (1562)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/16/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that an injection gold probe was used in the duodenum during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2017.According to the complainant, during the procedure, the distal tip of the injection gold probe fell off and was retrieved using a snare.The procedure was completed with another injection gold probe.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
|
|
Manufacturer Narrative
|
Investigation result: visual evaluation revealed that the ceramic distal tip was detached exposing the needle.Functional test was not performed due to the condition of the device.The complaint was confirmed.As specification from dfu, do not use the injection gold probe catheter through a duodenoscope.The elevator angle will damage the device.Based on all gathered information the device was used in a procedure that is not in accordance with the dfu.Therefore, the most probable root cause is use/user error.A review of the device history record (dhr) was performed and no deviations were found.A search of the complaint database confirmed that no other complaints exist for the specified batch.
|
|
Event Description
|
It was reported to boston scientific corporation that an injection gold probe was used in the duodenum during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2017.According to the complainant, during the procedure, the distal tip of the injection gold probe fell off and was retrieved using a snare.The procedure was completed with another injection gold probe.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
|
|