• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - COSTA RICA (COYOL) INJECTION GOLD PROBE¿; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - COSTA RICA (COYOL) INJECTION GOLD PROBE¿; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00560150
Device Problem Material Separation (1562)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/16/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that an injection gold probe was used in the duodenum during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2017.According to the complainant, during the procedure, the distal tip of the injection gold probe fell off and was retrieved using a snare.The procedure was completed with another injection gold probe.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Manufacturer Narrative
Investigation result: visual evaluation revealed that the ceramic distal tip was detached exposing the needle.Functional test was not performed due to the condition of the device.The complaint was confirmed.As specification from dfu, do not use the injection gold probe catheter through a duodenoscope.The elevator angle will damage the device.Based on all gathered information the device was used in a procedure that is not in accordance with the dfu.Therefore, the most probable root cause is use/user error.A review of the device history record (dhr) was performed and no deviations were found.A search of the complaint database confirmed that no other complaints exist for the specified batch.
 
Event Description
It was reported to boston scientific corporation that an injection gold probe was used in the duodenum during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2017.According to the complainant, during the procedure, the distal tip of the injection gold probe fell off and was retrieved using a snare.The procedure was completed with another injection gold probe.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INJECTION GOLD PROBE¿
Type of Device
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS  
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key6604951
MDR Text Key76431904
Report Number3005099803-2017-01598
Device Sequence Number1
Product Code KNS
UDI-Device Identifier08714729056492
UDI-Public(01)08714729056492(17)20180511(10)19238984
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K961349
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/11/2018
Device Model NumberM00560150
Device Catalogue Number6015
Device Lot Number19238984
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/13/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-