PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 SILICONE CHANNEL DRAIN WITHOUT TROCAR, 19 FR., ROUND HUBLESS, FULL FLUTED; DRAIN CHANNEL Back to Search Results
Catalog Number V072230
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "indications: bard® channel drains, round and flat silicone, are indicated for use with selected bard® evacuators for closed wound drainage following head and neck, orthopedic, abdominal, ent, ob/gyn, plastic, neurosurgery, thoracic and cardiovascular (channel drains only) procedures.Warnings: an effective closed suction drain system requires maintenance of the system to preserve patency.The drain must not be allowed to occlude nor the reservoir to completely fill; and reservoir suction must be maintained in order for the system to function properly.Verify that the system is functioning properly.If the system is not maintained properly, surgical complications, including hematomas, may result.In the event of occlusion of the drain, all wound drainage via the drain ceases.Careful attention to the drain will minimize the possibility of this problem.If occlusion does occur, the drain can be aspirated by connecting suction to the reservoir outlet or temporarily disconnecting the drain from the reservoir and applying suction directly to the drain.If an air-tight seal between the drain and the skin where the drain emerges is not achieved, the air leak must be rectified or the system must be converted to open drainage.An airtight seal between all system components (drain, adaptor and reservoir) is necessary for proper system function.Leaving the soft silicone elastomer drain implanted for any period of time so as to cause tissue ingrowth around the drain can interfere with easy removal and may effect the performance of the drain.The surgeon should monitor the patient¿s rate of wound healing.Evacuators should be used in cardio-thoracic surgery only after the lung is fully expanded and all air leaks have sealed.Drain perforations or channels must lie within the wound or cavity to be drained, otherwise inadequate drainage may result.To avoid the possibility of drain damage or breakage: · avoid suturing through drains.· drains should lie flat and in line with the skin exit areas.· particular care should be taken to avoid any obstacles to the drain exit path.· drains should be checked for free motion during closure to minimize the possibility of breakage.· drain removal should be done gently by hand.Drains should not be handled with pointed, toothed or sharp instruments which could cause cuts or nicks and lead to subsequent structural failure of the drain.· surgical removal may be necessary if drain is difficult to remove or breaks.Complications : complications which may result from the use of this suction drainage system include the risks associated with methods utilized in the surgical procedure, as well as the patients degree of intolerance to any foreign object in the body.The advantages of wound drainage, particularly closed system drainage, are lost if an air-tight seal between the drain and the skin where the drain emerges is not achieved, or if the drain is allowed to become occluded or if the reservoir is not activated properly, doesn¿t function properly or is not monitored.Evacuators should be emptied and re-activated when required per hospital protocol.In the event an air-tight seal is not achieved, the reservoir will rapidly fill with air from the leak; subsequent drainage to the reservoir will occur only if allowed by gravity and wound exudate forcing the flow.Entry into the reservoir is allowed only by displacement of air in the reservoir by wound exudate flow.In this displacement process, air reflux from the reservoir to the wound can occur and increase the likelihood of back-contamination across the anti-reflux valve.In the event of drain occlusion by fibrin, clots, or other particulate matter, all wound drainage via the drain ceases.If the reservoir is not emptied when it is full, equilibrium between the drain and reservoir at wound pressure will ultimately occur and drainage from the wound site will cease.When the reservoir and drain are at the same pressure and the reservoir is full of fluid, the likelihood of back-contamination across the anti-reflux valve is increased.When used to drain the pleural cavity in the presence of an air leak, drains must be attached to an appropriate pleural cavity drainage system to prevent tension pneumothorax.Drain placement · the surgeon should irrigate the wound with sterile fluid, then suction the irrigating fluid and gross debris from the operative site.· tubes should lie flat and in line with the anticipated skin exit.To facilitate later removal by manual traction, the tubing should not be curled, pinched, or sutured internally.· positioning of the drain in the body cavity, as well as the number of drains indicated, should be determined by the operating surgeon.· drain tubing should be placed within the wound by approximating the areas of critical fluid collection.· care must be taken to ensure that all drain perforations or channels lie completely within the wound or cavity to be drained.· taping or a triple loop suture (around and not through the tubing) will aid in preventing accidental drain displacement.· deep drainage is best accomplished by using one or more drains for each level of tissue.Each level should be evacuated by a separate vacuum source.· care must be exercised to avoid damage to the drain (see warnings).The tubing should be repeatedly checked during closure for free motion to avoid breakage and/or fragment retention within the wound.Do not use if package is damaged for use with 100cc, 200cc, and 400cc silicone evacuator, reliavac® 100cc evacuators, reliavac® 400cc evacuators, and 3-spring evacuators." (b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event or Problem Description
It was reported that the drain appeared to be the wrong size.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the drain appeared to be the wrong size.
 
Additional Manufacturer Narrative
No sample was returned for evaluation.However, it was identified to be an issue of mix with the lot number reported ngaz3014 (24fr with 19fr).Based on the investigation findings the reported problem is confirmed as manufacturing related.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "drain placement and the fr size is indicated: reorder #: drain size: drain shape: fluting trocar depth indicator: v072230, 19 fr.Round hubless, full ¿, dot.V072234, 24 fr.Round hubless, full ¿, dot.For use with 100cc, 200cc, 400cc, and 3-spring evacuators.Indications: round silicone channel drains are indicated for use with selected evacuators for closed wound drainage following head and neck, orthopedic, abdominal, ent, ob/gyn, plastic, neurosurgery, thoracic and cardiovascular (channel drains only) procedures.Complications: complications which may result from the use of this suction drainage system include the risks associated with methods utilized in the surgical procedure, as well as the patients degree of intolerance to any foreign object in the body.The advantages of wound drainage, particularly closed system drainage, are lost if an air-tight seal between the drain and the skin where the drain emerges is not achieved, or if the drain is allowed to become occluded or if the reservoir is not activated properly, doesn¿t function properly or is not monitored.Evacuators should be emptied and re-activated when required per hospital protocol.In the event an air-tight seal is not achieved, the reservoir will rapidly fill with air from the leak; subsequent drainage to the reservoir will occur only if allowed by gravity and wound exudate forcing the flow.Entry into the reservoir is allowed only by displacement of air in the reservoir by wound exudate flow.In this displacement process, air reflux from the reservoir to the wound can occur and increase the likelihood of back-contamination across the anti-reflux valve.In the event of drain occlusion by fibrin, clots, or other particulate matter, all wound drainage via the drain ceases.If the reservoir is not emptied when it is full, equilibrium between the drain and reservoir at wound pressure will ultimately occur and drainage from the wound site will cease.When the reservoir and drain are at the same pressure and the reservoir is full of fluid, the likelihood of back-contamination across the anti-reflux valve is increased.When used to drain the pleural cavity in the presence of an air leak, drains must be attached to an appropriate pleural cavity drainage system to prevent tension pneumothorax.Drain placement: ¿ the surgeon should irrigate the wound with sterile fluid, then suction the irrigating fluid and gross debris from the operative site.¿ tubes should lie flat and in line with the anticipated skin exit.To facilitate later removal by manual traction, the tubing should not be curled, pinched, or sutured internally.¿ positioning of the drain in the body cavity, as well as the number of drains indicated, should be determined by the operating surgeon.¿ drain tubing should be placed within the wound by approximating the areas of critical fluid collection.¿ care must be taken to ensure that all drain perforations or channels lie completely within the wound or cavity to be drained.¿ taping or a triple loop suture (around and not through the tubing) will aid in preventing accidental drain displacement.¿ deep drainage is best accomplished by using one or more drains for each level of tissue.Each level should be evacuated by a separate vacuum source.¿ care must be exercised to avoid damage to the drain (see warnings).The tubing should be repeatedly checked during closure for free motion to avoid breakage and/or fragment retention within the wound." (b)(4).The device was not returned.
 
Event or Problem Description
It was reported that the drain appeared to be the wrong size.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SILICONE CHANNEL DRAIN WITHOUT TROCAR, 19 FR., ROUND HUBLESS, FULL FLUTED
Common Device Name
DRAIN CHANNEL
Manufacturer (Section D)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX  85621
MDR Report Key6606023
Report Number1018233-2017-02850
Device Sequence Number16294181
Product Code GBX
UDI-Device Identifier00801741049903
UDI-Public(01)00801741049903(17)220131(10)NGAZ3014
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date01/31/2022
Device Catalogue NumberV072230
Device Lot NumberNGAZ3014
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/09/2017
Supplement Date Received by ManufacturerNot provided
07/12/2017
Initial Report FDA Received Date06/01/2017
Supplement Report FDA Received Date06/26/2017
07/12/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-