Model Number 112264-001 |
Device Problem
Hole In Material (1293)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/12/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).(b)(4).Device evaluated by mfr: it is indicated that the device will not be returned for evaluation; therefore a failure analysis of the complaint device could not be completed.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause was unable to be determined.(b)(4).
|
|
Event Description
|
It was reported that there was a hole in the catheter.A 2.1mm jetstream xc atherectomy catheter was selected for an atherectomy procedure in the superficial femoral artery.During the procedure, it was noted that the "clear coating" on the catheter shaft was raised for an area approximately an inch long but attached on both ends.It did not come off.Aspiration was in use but not negatively affected.During removal from the patient, the device was leaking at that spot as if a hole was in the catheter and saline was dripping out.The procedure was completed with a different device.There were no patient complications and the patient's condition was fine.
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was further reported that the lesion was heavily calcified and extremely difficult.There was difficulty in advancing the catheter.The raised "clear coating¿ on the catheter shaft was noticed after the withdrawal of the catheter from the patient.The case lasted 5 hours.
|
|
Search Alerts/Recalls
|