| Model Number |
R SERIES |
| Medical Device Problem Code |
Improper Alarm (2951)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a faulty solder joint on the pace/defib engine assembly.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device displayed a "defib disabled" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|