| Model Number |
LS3092 |
| Medical Device Problem Code |
Detachment Of Device Component (1104)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
04/18/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of initial report: (b)(6) 2017.The incident sample was requested but to date has not been received for evaluation.If the sample or additional information pertinent to the incident is obtained, a follow-up report will be submitted.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The customer reported that during assembly of the device, one of the prongs broke off.The device had not been used on a patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of initial report: (b)(6) 2017, date of follow-up report: (b)(6) 2017 one used ls3092 ligasure device was received, and evaluation of the sample found one of the snaps on the jaw was broken and missing.Snap damage can be caused by misaligning the snaps during assembly or by multiple assembly attempts.The investigation identified the root cause of the reported event to be improper assembly during use.The instructions included with this product lists measures for handling and assembling the device.If information is provided in the future, a supplemental report will be issued.
|
| |
|