• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS EMERALD; SURGICAL ADJUNCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS EMERALD; SURGICAL ADJUNCTS Back to Search Results
Model Number EMERALDC30
Device Problem Material Distortion (2977)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/08/2017
Event Type  malfunction  
Manufacturer Narrative
This is not an implantable device.All pertinent information available to abbott medical optics has been submitted.
 
Event Description
It was reported that the cartridge tip was noticed distorted before the insertion of the intraocular lens (iol).No patient injury was reported.No further information was provided.Two cartridge tips were reported distorted; therefore two mdrs will be filed.This report is for 1 of 2.
 
Manufacturer Narrative
Device evaluation.Only unused cartridges of the same lot number were returned as representative samples.All 59 returned emeraldc30 cartridges were inspected at 10x microscope magnification and no cracks at the tip were observed.All units were observed in good condition.The customer's reported complaint was not verified.Manufacturing records review: the manufacturing records for the cartridge were reviewed.During the manufacturing process the operators check the neck, tube and tip areas for cracks.No cracking or stress marks are allowed.They also check the tip for any melting, roughness, dent, bent tip or smash condition.The product was manufactured and released according to specification.A search revealed that no additional investigation requests for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERALD
Type of Device
SURGICAL ADJUNCTS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
Manufacturer (Section G)
ABBOTT MEDICAL OPTICS INC.
road 402 north, km 4.2
anasco industrial park, pob 14
anasco PR 00610
Manufacturer Contact
pam mcclain
1700 east st. andrew place
santa ana, CA 92705
7142478243
MDR Report Key6607312
MDR Text Key76502616
Report Number2648035-2017-00977
Device Sequence Number1
Product Code KYB
UDI-Device Identifier05050474530027
UDI-Public(01)05050474530027(17)171004(10)CB40353
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K961242
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/02/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date10/04/2017
Device Model NumberEMERALDC30
Device Catalogue NumberEMERALDC30
Device Lot NumberCB40353
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/02/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/04/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-