| Model Number |
180733-2 |
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/08/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Poly failed to fit into baseplate so we had to open a new one and the new one went right in.It seems to be just misshaped.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding seating/locking issues involving a mako insert was reported.The event was not confirmed.However, the material analysis report was able to confirm damage.Method & results: device evaluation and results: examination of the device by the material analysis engineer indicated that damage observed on distal surface consistent with contact against a hard object.Medical records received and evaluation: not performed as no medical records were provided.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies complaint history review: there has been no other event for the lot referenced.Conclusions: the event reported for poly failed to fit into baseplate but it is confirmed through visual inspection and material analysis engineer indicated that damage observed on distal surface consistent with contact against a hard object.If the additional information are received, this investigation will be reopened and re-evaluated.No further investigation for this event is required at this time.
|
| |
|
Event or Problem Description
|
|
Poly failed to fit into baseplate so we had to open a new one and the new one went right in.It seems to be just misshaped.
|
| |
|
Search Alerts/Recalls
|