NEWPORT MEDICAL INSTRUMENTS INC. E360 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number E360
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the 360 ventilator air supply was lost.There was no patient involvement at the time of the reported condition.Covidien was not authorized to evaluate/service the device so the reported complaint could not be confirmed.A non-covidien service provider checked the ventilator, and confirmed the customer issue.The service provider tightened all internal filters, performed flow sensor and oxygen sensor calibrations.The ventilator was tested and checked to be running normally.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E360 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
NEWPORT MEDICAL INSTRUMENTS INC.
1620 sunflower ave
costa mesa CA 92626
MDR Report Key6608281
Report Number2023050-2017-05375
Device Sequence Number1097141
Product Code CBK
UDI-Device Identifier10884521540989
UDI-Public10884521540989
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K053502
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE360
Device Catalogue NumberE360E-WW-IN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2017
Initial Report FDA Received Date06/02/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-