Catalog Number C-HSK-3038 |
Device Problem
Failure to Fire (2610)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/05/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).A lot history record review was completed for the reported product lot number.There was no nonconformance recorded in the lot history.The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
|
|
Event Description
|
The hospital reported that during a coronary artery bypass procedure, hs iii proximal seal system 3.8mm did not fire.A replacement device was used to complete the procedure.The hospital did not report any patient effects.
|
|
Manufacturer Narrative
|
(b)(4).The hsk device was returned to the factory for evaluation.Signs of clinical use and evidence of blood were observed.Specks of blood was observed on the loading device.The delivery device was returned inside the loading device.A visual inspection was conducted.The delivery device was removed from the loading device.The seal and tension spring assembly remained in the loading device.The slide lock was engaged.The plunger was not depressed on the delivery device.The seal was taken out from the loading device for inspection.No crack/delamination of seal was observed.The following measurements were taken; the inner delivery tube diameter was measured at.199 in.The outer diameter was measured at.221 in.The length of the delivery tube was measured at 2.51 in.The values recorded were within the tolerance specifications.Based upon the received condition of the device, the reported complaint for "failure to deploy" was not confirmed and the analyzed failure "fitting problem", was confirmed.
|
|
Event Description
|
The hospital reported that during a coronary artery bypass procedure, hs iii proximal seal system 3.8mm did not fire.The hospital did not report any patient effects.
|
|