MAUDE Adverse Event Report: MAQUET CV HS III PROXIMAL SEAL SYTEM 3.8MM; CLAMP, VASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CV HS III PROXIMAL SEAL SYTEM 3.8MM; CLAMP, VASCULAR Back to Search Results
Catalog Number C-HSK-3038
Device Problem Failure to Fire (2610)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/05/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A lot history record review was completed for the reported product lot number.There was no nonconformance recorded in the lot history.The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Event Description
The hospital reported that during a coronary artery bypass procedure, hs iii proximal seal system 3.8mm did not fire.A replacement device was used to complete the procedure.The hospital did not report any patient effects.
 
Manufacturer Narrative
(b)(4).The hsk device was returned to the factory for evaluation.Signs of clinical use and evidence of blood were observed.Specks of blood was observed on the loading device.The delivery device was returned inside the loading device.A visual inspection was conducted.The delivery device was removed from the loading device.The seal and tension spring assembly remained in the loading device.The slide lock was engaged.The plunger was not depressed on the delivery device.The seal was taken out from the loading device for inspection.No crack/delamination of seal was observed.The following measurements were taken; the inner delivery tube diameter was measured at.199 in.The outer diameter was measured at.221 in.The length of the delivery tube was measured at 2.51 in.The values recorded were within the tolerance specifications.Based upon the received condition of the device, the reported complaint for "failure to deploy" was not confirmed and the analyzed failure "fitting problem", was confirmed.
 
Event Description
The hospital reported that during a coronary artery bypass procedure, hs iii proximal seal system 3.8mm did not fire.The hospital did not report any patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HS III PROXIMAL SEAL SYTEM 3.8MM
Type of Device
CLAMP, VASCULAR
Manufacturer (Section D)
MAQUET CV
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
MAQUET CV
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
45 barbour pond drive
wayne, NJ 07470
MDR Report Key6610163
MDR Text Key76727206
Report Number2242352-2017-00541
Device Sequence Number1
Product Code DXC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080169
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/03/2017
Device Catalogue NumberC-HSK-3038
Device Lot Number25129292
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2017
Is the Reporter a Health Professional? No
Device AgeYR
Date Manufacturer Received06/27/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/03/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age68 YR
-
-