• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CORPORATION POWERFLEXPRO 5MM15CM 135; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION POWERFLEXPRO 5MM15CM 135; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 4400515X
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/09/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The products were not returned for analysis.A review of the manufacturing records could not be conducted without a lot number.Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
It was reported that five (5) powerflex pro balloon catheter (bc) burst at ten to twelve atmosphere (10-12 atm) during the intervention in the same patient.There was no patient injury.The device was prep normally.There were no difficulty inflating the balloons during prep. there were no difficulties removing the product from the hoop, the protective balloon cover, the stylet or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting the product the product into the patient.The product was removed intact (in one piece) from the patient.Contrast media used was solutrast 350.The contrast to saline ratio was cm+nacl.The type and brand of inflation device used was arcroyal which was used successfully with other devices.There were no resistance/friction noted while inserting the balloon through the rotating hemostatic valve, guide catheter, and through the vessel.There were no difficulty crossing the lesion.
 
Additional Manufacturer Narrative
This is one of five products involved with the reported event.The manufacturing reference number for this event is(b)(4).The manufacturing reference number for the other complaints are (b)(4).  it was reported that five (5) powerflex pro balloon catheter (bc) burst at ten to twelve atmosphere (10-12 atm) during the intervention in the same patient.There was no patient injury.The lesion was described as massively calcified.The device was prep normally.There were no difficulty inflating the balloons during prep.There were no difficulties removing the product from the hoop, the protective balloon cover, the stylet or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting the product the product into the patient.The product was removed intact (in one piece) from the patient.Contrast media used was solutrast 350.The contrast to saline ratio was cm and nacl.The type and brand of inflation device used was arcroyal which was used successfully with other devices.There were no resistance/friction noted while inserting the balloon through the rotating hemostatic valve, guide catheter, and through the vessel.There were no difficulty crossing the lesion.The device was not returned for analysis.A device history record (dhr) review could not be conducted as a lot number was not provided.The reported ¿balloon burst-at/below rbp¿ could not be confirmed as the devices were not returned for analysis.The exact cause of the bursts reported could not be determined.Based on the limited information available for review, vessel characteristics (massive calcification) may have contributed to the reported event as calcification/resistant lesions may cause damage to the balloon.According to the instructions for use, which is not intended as a mitigation, ¿the rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ given the limited information available for review at this time, there is nothing to suggest that the reported event is related to the design and manufacturing process of the device; therefore no corrective action will be taken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERFLEXPRO 5MM15CM 135
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60th ave
miami lakes FL 33014
MDR Report Key6610570
Report Number9616099-2017-01139
Device Sequence Number10764875
Product Code LIT
UDI-Device Identifier20705032059432
UDI-Public20705032059432
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K112797
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Medical Assistant
Type of Report Initial,Followup
Report Date (Section B) 07/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4400515X
Device Catalogue Number4400515X
Was Device Available for Evaluation? No
Distributor Facility Aware Date05/09/2017
Initial Date Received by Manufacturer 05/09/2017
Supplement Date Received by Manufacturer07/03/2017
Initial Report FDA Received Date06/05/2017
Supplement Report FDA Received Date07/06/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-