Catalog Number 72200082
Device Problems Peeled/Delaminated (1454); Metal Shedding Debris (1804)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/11/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the burrs were shedding and flaking.Debris was removed from the patient.A backup device was available to complete the procedure.No patient impact was reported.
 
Manufacturer Narrative
Examination was not possible, as the device was not returned.The investigation was limited to the information provided.The investigation could not draw any conclusions about the reported event with the clinical details provided.A review of the device history record was performed which confirmed no inconsistencies.No further investigation is warranted at this time.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BURR,5.5MM ABRADER,180MML DISPOSABLE (3
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer Contact
james gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key6611517
MDR Text Key76730445
Report Number1219602-2017-00581
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/15/2019
Device Catalogue Number72200082
Device Lot Number50647284
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received07/20/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/15/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-