| Model Number |
26120BA |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
05/05/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Scope has not been returned for evaluation.
|
| |
|
Event or Problem Description
|
|
Allegedly, the doctor was performing a d & c hysteroscopy with ablation on patient, when she removed scope from uterus and noticed a piece of the scope distal window was cracked and missing.They brought in another scope and completed procedure.Doctor went back in to check if piece was inside patient.She didn't find it; no x-ray was performed.Doctor doesn't have plans to schedule another procedure to locate missing piece.She believes that it was flushed out during irrigation.The hospital reported there was no injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The distal window seal is broken.Possible causes can be the epoxy coming undone on distal lens window.Excessive or prolonged heat will cause this along with trauma to the instrument distal tip.We cannot confirm cause.
|
| |
|
Event or Problem Description
|
|
This is a supplemental.I will be adding method/results/conclusion codes and evaluation of the instrument.
|
| |
|