MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R Back to Search Results
Model Number 5DCS
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/05/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon visual inspection, the device showed signs of clinical use.Debris was present on the blades of the device.Inspection of the blades revealed wear.Functional inspection revealed the blades were dull and did not provide a clean cut due to the wear present on the blades.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root cause(s) are: - user attempts to cut staples, clips, or non-soft tissue.- wear to blades due to clinical use.The instructions for use (ifu) state: - do not directly apply current to staples or clips.- instruments were designed for cutting soft tissue.Attempting to cut staples or clips may damage the instrument.- if cutting of staples or clips is attempted, damage to instrument may occur.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported the laparoscopic scissors were not cutting.There was a five minute delay to grab a new device.There was no patient injury or medical intervention reported.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5307 great oak drive
lakeland FL 33815
MDR Report Key6613591
Report Number0001056128-2017-00078
Device Sequence Number602249
Product Code NUJ
UDI-Device Identifier00885825013288
UDI-Public(01)00885825013288(17)200315(10)5975643
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K110189
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/15/2020
Device Model Number5DCS
Device Catalogue Number5DCSRR
Device Lot Number5975643
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2017
Initial Report FDA Received Date06/05/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Patient Age82 YR
-
-