• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE 14F; REPLACEMENT G-TUBES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE 14F; REPLACEMENT G-TUBES Back to Search Results
Model Number N/A
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/05/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer, at this time, for evaluation.A lot history review (lhr) of ngav1538 showed no other similar product complaint(s) from this lot number.Device not yet returned for evaluation.
 
Event or Problem Description
It was reported that the device was placed on (b)(6) 2017 and removed on (b)(6) 2017 due to the cuff on the device breaking.No other information was provided.No harm to the patient.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a damaged tri-funnel inner balloon bolster was confirmed and the cause appears to be use related.The product returned for evaluation was a 14fr tri-funnel feeding tube.The damage observed in the returned sample was characteristic of over-pressurization (burst) damage.This can occur through over-filling the feeding tube balloon.The appropriate inflation volume is listed on the inflation valve and over-inflation can be avoided by fully draining the balloon prior to inflation.The returned product sample was evaluated and a longitudinally-aligned split in the balloon membrane was observed.This catheter damage was determined to be typical of a burst, and the characteristics observed which supported this type of failure included: tensile weakness in the membrane at the fracture site (due to over-inflation prior to burst.Granular fracture surface texture (typical of tearing failure modes).Additionally, residual material and discoloration was seen throughout the sample which established a usage history of the product and suggested the product had worked as intended for a period of time prior to the event.An examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.These features are characteristically found due to the over-inflation and fracture of the balloon tubing.The product instruction for use (ifu) contains the following information which may be relevant to this type of event, "do not exceed the maximum recommended inflation volume as this may cause excessive pressure on the gastric mucosa and migration of the internal balloon into the peritoneal cavity".
 
Event or Problem Description
It was reported that the device was placed on (b)(6) 2017 and removed on (b)(6) 2017 due to the cuff on the device breaking.No other information was provided.No harm to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE 14F
Common Device Name
REPLACEMENT G-TUBES
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key6616779
Report Number3006260740-2017-00727
Device Sequence Number13637947
Product Code KNT
UDI-Device Identifier00801741037122
UDI-Public(01)00801741037122(17)190331(10)NGAV1538
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K063118
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 08/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2019
Device Model NumberN/A
Device Catalogue Number000714
Device Lot NumberNGAV1538
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2017
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer08/01/2017
Initial Report FDA Received Date06/06/2017
Supplement Report FDA Received Date08/09/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age24 YR
-
-