• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS SENSAR; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS SENSAR; MONOFOCAL IOLS Back to Search Results
Model Number AR40E
Medical Device Problem Code Detachment Of Device Component (1104)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 05/12/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Initial reporter phone number: (b)(6).Implant and explant dates: if explanted; give date: not applicable as the lens was inserted and removed.(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that during the implantation of an intraocular lens (iol) into the right eye of a (b)(6) year-old male patient, one of the haptics was detached.Reportedly, the incision had to be enlarged to remove and replace the lens.The incision was closed with 3 points.The patient outcome was reported being good.No further information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the intraocular lens (iol) was not returned at the manufacturing site; therefore product testing could not be performed and the customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no additional complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSAR
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6616820
Report Number2648035-2017-01008
Device Sequence Number979339
Product Code HQL
UDI-Device Identifier05050474502055
UDI-Public(01)05050474502055(17)211222
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date12/22/2021
Device Model NumberAR40E
Device Catalogue NumberAR40E00175
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/12/2017
Supplement Date Received by Manufacturer07/10/2017
Initial Report FDA Received Date06/06/2017
Supplement Report FDA Received Date08/02/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
-
-