|
A healthcare professional (hcp) via a manufacturer representative reported introperative test electrode group of ¿anti-high,¿ repeated connection unresolved.Factors noted to have led to the event were noted as ¿causes of implanted product/patient edema during surgery.¿ troubleshooting was noted as intraoperative multiple tests, impedance was still high on all electrode combinations (>4000 ohms), excluding the reasons for connection.Suspected impedance itself high, replace the electrode, test again.It was noted that during the surgery, replaced with a new electrode, and the impedance test was normal.The problem had resolved.No symptoms were reported.There were no further complications reported and/or anticipated.
|