| Model Number |
H74939293068030 |
| Medical Device Problem Code |
Premature Activation (1484)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/16/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that inadvertent stent deployment occurred.During preparation for a stenting procedure, the physician selected an innova 6 x 80 x 130 stent.The stent deployed outside the patient on the table.The device never entered the patients body.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Returned product consisted of an innova self-expanding stent delivery system (sds).The outer sheath, middle shaft and the remainder of the device were checked for damage.The outer sheath showed a kink at the nosecone.There was also a kink approximately 43cm from the nosecone on the outer sheath.No additional damage was noticed.The stent was returned inside the device and it was inadvertently partially deployed approximately 8mm from the marker band.Ops engineering opened the handle to inspect for additional damage, no additional damage was noticed.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause was unable to be determined.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that inadvertent stent deployment occurred.During preparation for a stenting procedure, the physician selected an innova 6 x 80 x 130 stent.The stent deployed outside the patient on the table.The device never entered the patients body.No patient complications were reported.
|
| |
|
Search Alerts/Recalls
|