MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE INNOVA¿; STENT, SUPERFICIAL FEMORAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE INNOVA¿; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Model Number H74939293068030
Medical Device Problem Code Premature Activation (1484)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 05/16/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that inadvertent stent deployment occurred.During preparation for a stenting procedure, the physician selected an innova 6 x 80 x 130 stent.The stent deployed outside the patient on the table.The device never entered the patients body.No patient complications were reported.
 
Additional Manufacturer Narrative
Returned product consisted of an innova self-expanding stent delivery system (sds).The outer sheath, middle shaft and the remainder of the device were checked for damage.The outer sheath showed a kink at the nosecone.There was also a kink approximately 43cm from the nosecone on the outer sheath.No additional damage was noticed.The stent was returned inside the device and it was inadvertently partially deployed approximately 8mm from the marker band.Ops engineering opened the handle to inspect for additional damage, no additional damage was noticed.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause was unable to be determined.(b)(4).
 
Event or Problem Description
It was reported that inadvertent stent deployment occurred.During preparation for a stenting procedure, the physician selected an innova 6 x 80 x 130 stent.The stent deployed outside the patient on the table.The device never entered the patients body.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOVA¿
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
MDR Report Key6619501
Report Number2134265-2017-05499
Device Sequence Number11762
Product Code NIP
UDI-Device Identifier08714729874010
UDI-Public(01)08714729874010(17)20200229(10)20347587
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P140028
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2020
Device Model NumberH74939293068030
Device Catalogue Number39293-06803
Device Lot Number20347587
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 05/16/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date06/07/2017
Supplement Report FDA Received Date06/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-