Model Number ETBF3620C145E
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Codes Pain (1994); Rupture (2208)
Date of Event 05/15/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An endurant ii stent graft system was implanted in a patient for the endovascular treatment of a 60mm diameter abdominal aortic aneurysm.It was reported that the patient presented emergently with back pain.A ct was done and showed there was a ia endoleak and rupture.The physician stated when the stent graft was originally placed they used aptus and placed several endoanchors because the patient had a short and large diameter aortic neck.The stent graft remained where it was originally placed but disease progression moved upward to the level of the renal arteries.The physician snorkeled the right renal and extended using an endurant cuff that was placed just below the level of the sma.The physician stated the cause of the event was anatomy related.No additional clinical sequelae were reported and the patient is fine.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II BIFURCATED STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key6619774
Report Number2953200-2017-00937
Device Sequence Number738827
Product Code MIH
UDI-Device Identifier00613994991058
UDI-Public00613994991058
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/06/2014
Device Model NumberETBF3620C145E
Device Catalogue NumberETBF3620C145E
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/15/2017
Supplement Date Received by Manufacturer05/15/2017
Initial Report FDA Received Date06/07/2017
Supplement Report FDA Received Date10/04/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
-
-