Catalog Number 03.113.025
Medical Device Problem Code Component Missing (2306)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There was no known reported patient involvement associated with the complained event.Unknown when device malfunctioned.Device is an instrument and is not implanted/explanted.A device history record (dhr) review was performed for part # 03.113.025, lot # 3665631: manufacturing site:(b)(4), manufacturing date: 28.Jan.2011: no non conformance reports (ncrs) were generated during production.Review of the device history record showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Service history review was performed for part # 03.113.025, lot # 3665631: no service history review can be performed as part number 03.113.025 with lot number(s) 3665631 is a lot/batch controlled item.The manufacture date of this item is 1-feb-2011.The source of the manufacture date is the release to warehouse date.The service history review is unconfirmed.The device has been received and the product evaluation is in progress.No conclusion can be drawn.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that an insertion screw for an insertion handle is missing.No patient involvement reported.This report is for one (1) insertion handle.This is report 1 of 1 for complaint com-(b)(4).
 
Additional Manufacturer Narrative
A service and repair evaluation was performed on subject device.The customer reported the insertion screw was missing.The repair technician reported the screw was missing.Missing parts is the reason for repair.The cause of the issue is unknown.The following parts were replaced: m3.5 screw.The item was repaired per the inspection sheet, passed synthes final inspection on 1-jun-2017 and will be returned to the customer upon completion of the service and repair process.The evaluation was confirmed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSERTN HANDLE F/3.5MM LOW BNDMEDIAL DISTAL TIB PLATE/RIGHT
Common Device Name
GAUGE,DEPTH
Manufacturer (Section D)
SYNTHES HAGENDORF
im bifang 6
hagendorf CH461 4
SZ  CH4614
MDR Report Key6619838
Report Number3003875359-2017-10274
Device Sequence Number16166906
Product Code HTJ
UDI-Device Identifier10886982072917
UDI-Public(01)10886982072917(10)3665631
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.113.025
Device Lot Number3665631
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 05/12/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date06/07/2017
Supplement Report FDA Received Date06/22/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-