| Model Number |
HGK1809 |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
Blood Loss (2597)
|
| Date of Event |
04/18/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the complaint device history records, indicated that the graft was processed and inspected according to procedures and was therefore released following acceptable quality inspections and tests.Specifically, no anomaly was evidenced in the collagen coating records.The review of the water permeability testing records of products coated on the same day as the complaint device indicated values well within product specifications.Please note that water permeability testing is recognized by international standard for vascular grafts as the test to address the risk of blood leakage.One retention sample coated on the same day and under the same conditions as the involved device underwent water permeability testing.The test result indicated a value well within product specifications (< 5 ml/cm²/min).No conclusion can be drawn.However, the conducted investigation and testing performed would tend to indicate that the product was not defective.
|
| |
|
Event or Problem Description
|
|
During an aorto-bifemoral procedure, performed on (b)(6) 2017, it was reported there was a lot of bleeding during the procedure.The bleeding was stopped by clamping the graft.The patient had a low blood pressure, and unexpected blood transfusion was performed.It was also reported that before the aorto-bifemoral implantation, a temporary axillobifemoral shunt was performed with another graft hgk0010-100 without bleeding.No consequence for the patient was reported.
|
| |
|
Search Alerts/Recalls
|