|
The insulin pump was received with all operating currents within spec and passed functional testing including the rewind, basic occlusion test, occlusion test, prime test, excessive no delivery test, self test, unexpected restart error test, delivery accuracy test and displacement test.The insulin pump delivered properly all bolus programmed during testing.The insulin pump was monitored for 2 days and no unexpected delivery of bolus anomaly noted.The insulin pump was received with cracked case at the display window corners, display window, reservoir tube, reservoir tube lip, battery tube threads and belt clip slot.Unit received with minor scratched display window and case.A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.
|