Model Number 6227DEF04
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/27/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that while slitting the delivery catheter during implant, the physician got snagged at the distal portion of the catheter and it pulled the left ventricular (lv) lead from the coronary sinus.It was noted the snag appeared to be on the wire for deflection portion.The delivery catheter was replaced to gain cs access a second time and the lv lead remains in use.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN PREVAIL DEFLECTABLE CATHETER DELIVERY SYSTEM
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key6626788
Report Number9612164-2017-00703
Device Sequence Number18854701
Product Code DQY
UDI-Device Identifier00613994450753
UDI-Public00613994450753
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K061480
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/07/2018
Device Model Number6227DEF04
Device Catalogue Number6227DEF04
Device Lot Number0008405314
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/27/2017
Supplement Date Received by Manufacturer03/27/2017
Initial Report FDA Received Date06/08/2017
Supplement Report FDA Received Date10/04/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age83 YR
Patient Weight45
-
-