| Model Number |
439678 |
| Medical Device Problem Codes |
Failure to Capture (1081); High Capture Threshold (3266); Unstable Capture Threshold (3269)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/06/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the left ventricular lead had high threshold.The lead was explanted and during the replacement procedure, two leads were attempted and not used.Both had no capture, and high and unstable threshold.Another lead was implanted.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the full lead was returned and analyzed.No anomalies were found.If information is provided in the future, a supplemental report will be issued.
|
| |
|