| Model Number |
6944-65 |
| Medical Device Problem Codes |
Over-Sensing (1438); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Chest Pain (1776); Shock from Patient Lead(s) (3162)
|
| Date of Event |
04/28/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a lead integrity alert (lia) triggered and the patient was admitted to the emergency room with chest pain due to inappropriate shocks by the right ventricular (rv) lead.The right ventricular (rv) lead had dislodged and was found to have oversensing with high rate non-sustained (hr-ns) and sensing integrity counter (sic) episodes.The lead was attempted to be repositioned but would not stay in place.The lead was removed and a new lead was implanted.It was further reported that the right atrial (ra) lead had dislodged.The lead was repositioned and remains in use.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the full lead was returned and analyzed.Analysis was performed and no anomalies were found.Visual analysis of the lead indicated apparent explant damage.If information is provided in the future, a supplemental report will be issued.
|
| |
|