| Model Number |
DTBB1D4 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Hematoma (1884); Shock from Patient Lead(s) (3162)
|
| Date of Event |
04/11/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant products: 6935m62 lead, implanted: (b)(6) 2017.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had an implantable cardioverter defibrillator (icd) pocket revision for hematoma.During the icd pocket revision, the right ventricular (rv) lead had noise oversensing during electrocautery use and delivered unexpected shock therapies to the patient.The icd and rv lead remains in use.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|