• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MARTECH MEDICAL PRODUCTS SCHON XL DOUBLELUMEN CATHETER; SCHON¿ HEMODIALYSIS DIALYSIS CATHETER SETS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MARTECH MEDICAL PRODUCTS SCHON XL DOUBLELUMEN CATHETER; SCHON¿ HEMODIALYSIS DIALYSIS CATHETER SETS Back to Search Results
Catalog Number H787108017015
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/16/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Angiodynamics notified the dialysis catheter manufacturer of this complaint event on (b)(6) 2016 via supplier corrective action request (b)(4).As part of a review of complaints related to distribution only products, angiodynamics identified that it does not have objective evidence that the manufacturer assessed this complaint event for mdr reportability.As a result of this review, angiodynamics chose to assess this complaint event for reportability and determined that it does meet the criteria to file an mdr.This retrospective mdr is being filed based on this review and to ensure complaint file is complete.One schon dialysis was returned for evaluation and forwarded to the vendor for evaluation.Per (b)(4) vendor response, pinholes were noted in both walls of the venous extension tubing.The customer's reported complaint of a leak in the tube is confirmed.As stated by the manufacturer, the most likely root cause for the reported complaint description for the holes on the tubing was due to the clinical procedure.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.The incoming receiver lot record and the vendor manufacturing lot records both indicate that all device specification and quality requirements were met.The instructions for use, which is supplied to the end user with this catalog number, states: "clamping the catheter repeatedly in the same spot could weaken the tubing.Change the position of the clamps regularly to prolong the life of the tubing." a review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.(b)(4).
 
Event or Problem Description
As reported to angiodynamics on (b)(6) 2016: temporary dialysis catheter was placed on patient.Blue port had a leak in it on the clamp portion of the line.The physician assistant was able to access the catheter and patient.Patient was able to complete their apheresis with the leak.The clamp may have caused a puncture in the line.The reported defective device has been returned to the manufacturer for an investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCHON XL DOUBLELUMEN CATHETER
Common Device Name
SCHON¿ HEMODIALYSIS DIALYSIS CATHETER SETS
Manufacturer (Section D)
MARTECH MEDICAL PRODUCTS
1500 delp drive
harleysville PA 19438
MDR Report Key6629740
Report Number1319211-2017-00062
Device Sequence Number1081994
Product Code MSD
UDI-Device IdentifierH787108017015
UDI-PublicH787108017015
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K971925
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician Assistant
Device Expiration Date01/31/2021
Device Catalogue NumberH787108017015
Device Lot NumberMHKA480
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2016
Initial Report FDA Received Date06/09/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
-
-