| Catalog Number |
9734403 |
| Medical Device Problem Code |
Bent (1059)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/12/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2017 a replacement probe was shipped.Replacement probe resolved the issue.No further issues were reported.The suspected probe was returned to manufacturer for evaluation.Testing found that the tip of the probe was twisted severely.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
| |
|
Event or Problem Description
|
|
A medtronic representative reported that during a spinal fusion procedure with navigation system a navlock thoracic probe was found to have bent tip.There was no delay to the procedure.The procedure was completed with the use of navigation system.No impact on patient outcome.
|
| |
|