Catalog Number 9734403
Medical Device Problem Code Bent (1059)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/12/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
On (b)(6) 2017 a replacement probe was shipped.Replacement probe resolved the issue.No further issues were reported.The suspected probe was returned to manufacturer for evaluation.Testing found that the tip of the probe was twisted severely.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Event or Problem Description
A medtronic representative reported that during a spinal fusion procedure with navigation system a navlock thoracic probe was found to have bent tip.There was no delay to the procedure.The procedure was completed with the use of navigation system.No impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THORACIC PROBE TIP
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
MDR Report Key6630225
Report Number1723170-2017-02415
Device Sequence Number16470291
Product Code OLO
UDI-Device Identifier00613994724014
UDI-Public00613994724014
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K124004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date (Section B) 06/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Catalogue Number9734403
Device Lot Number111102
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2017
Initial Report FDA Received Date06/09/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-