| Catalog Number |
9734723 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/08/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The suspect clamp was returned to the manufacturer for evaluation.Visual inspection found that the retainer ring was removed from the adjustment screw.This results in the clamp being able to tighten but not open.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
| |
|
Event or Problem Description
|
|
A medtronic representative reported that, while in a spinal fusion, the spine clamp washer fell off of the clamp.The washer was retrieved and the site elected to continue the procedure with a different clamp.There was a reported delay to the procedure of less than 1 hour due to this issue.There was no impact on patient outcome.No additional information was provided.
|
| |
|